Comparison of the Consistency and Difference of Three Detection Methods in Blood Concentration Monitoring of Mycophenolic Acid
- VernacularTitle:3种方法检测霉酚酸血药浓度的一致性与差异性比较
- Author:
Fengjun LYU
1
;
Si LI
1
;
Fengjun SUN
2
;
Jishan FAN
1
Author Information
1. Dept. of Pharmacy,Children’s Hospital of Chongqing Medical University/National Clinical Research Center for Child Health and Disorders/Ministry of Education Key Laboratory of Child Development and Disorders/Chongqing Key Laboratory of Pediatrics,Chongqing 400014,China
2. Dept. of Pharmacy,the First Affiliated Hospital of Army Medical University,Chongqing 400038,China
- Publication Type:Journal Article
- Keywords:
Mycophenolic acid;
Blood concentration;
Fluorescence immunochromatography;
Liquid chromatography-tandem
- From:
China Pharmacy
2021;32(23):2885-2889
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE:To investigate the consistency and difference of f luorescence immunochromatographic and liquid chromatography-tandem mass spectrometry (LC-MS/MS)and enzyme multiplied immunoassay technique (EMIT)in the blood concentration monitoring of mycophenolic acid. METHODS :Fluorescence immunochromatography ,LC-MS/MS and EMIT were used to detect the blood concentration of mycophenolic acid in 61 blood samples of children treated with mycophenolate mofetil ester orally at different time points. Kolmogorov-Smirnov method ,Wilcoxon pairing test ,Passing-Bablok regression ,Cusum method,Spearman correlation analysis ,Bland-Altman scatter diagram were adopted for statistical analysis. RESULTS :Blood concentrations of mycophenolic acid ,which were determined by fluorescence immunochromatography ,LC-MS/MS and EMIT , showed non-normal distribution. Passing-Bablok regression analysis showed that regression equation of fluorescence immunochromatography and LC-MS/MS ,fluorescence immunochromatographic method and EMIT were CFI=0.928 3CLC-MS/MS+0.961 7 and CFI=0.880 7CEMIT-0.488 2(FI means fluorescence immunochromatographic ). Spearman correlation analysis showed that the correlation coefficients between fluorescence immunochromatography and LC-MS/MS ,fluorescence immunochromatography and EMIT were 0.968 and 0.929, respectively (P<0.000 1). Bland Altman scatter plot analysis showed that 3.28% of the 358341451@qq.com difference between fluorescence immunochromatography and LC-MS/MS was outside the consistency limit (±1.96SD), and 1.64% of the difference between fluorescence immuno- chromatography and EMIT was outside the consistency limit (± 1.96SD). Wilcoxon pairing test showed that the results of fluorescence immunochromatography were higher than those of LC-MS/MS (Z=3.76,P=0.000 2)and lower than those of EMIT (Z=-5.96,P<0.000 1). CONCLUSIONS :Fluorescence immunochromatography shows good consistency and correlation with LC-MS/MS and EMIT ;the blood concentrations of mycophenolic acid detected by fluorescence immunochromatography were higher than those by LC-MS/MS and lower than those by EMIT . It can be used for bedside rapid detection. When using the test results of different methods for clinical medication ,the differences of test methods need to be considered.