Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils.
10.4168/aair.2015.7.4.404
- Author:
Sehyo YUNE
1
;
Jin Young LEE
;
Dong Chull CHOI
;
Byung Jae LEE
Author Information
1. Department of Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea. bj65.lee@samsung.com
- Publication Type:Brief Communication
- Keywords:
Asthma;
nitric oxide;
electrochemical technique;
eosinophils;
sputum
- MeSH:
Asthma;
Body Mass Index;
Electrochemical Techniques;
Eosinophils*;
Methacholine Chloride;
Nitric Oxide*;
ROC Curve;
Smoke;
Smoking;
Sputum*
- From:Allergy, Asthma & Immunology Research
2015;7(4):404-408
- CountryRepublic of Korea
- Language:English
-
Abstract:
This study was performed to compare the 2 different portable devices measuring fractional exhaled nitric oxide (FeNO) and to see the correlation between FeNO and induced sputum eosinophil count (ISE). Forty consecutive subjects clinically suspected to have asthma underwent FeNO measurement by NIOX-MINO(R) and NObreath(R) concurrently. All also had induced sputum analysis, methacholine provocation test or bronchodilator response test, and spin prick test. Agreement between the 2 devices was evaluated. The correlation between FeNO and ISE was assessed, as well as the cut-off level of FeNO to identify ISE > or =3%. The intraclass correlation coefficient (ICC) between FeNO levels measured by NIOX-MINO(R) (FeNO(NIOX-MINO)) and NObreath(R) (FeNO(NObreath)) was 0.972 with 95% confidence interval of 0.948-0.985. The 95% limits of agreement were -28.9 to 19.9 ppb. The correlation coefficient between ISE and FeNO(NIOX-MINO) was 0.733 (P<0.001), and 0.751 between ISE and FeNO(NObreath) (P<0.001). The ROC curve found that the FeNO(NIOXMINO) of 37.5 ppb and the FeNO(NObreath) of 36.5 ppb identified ISE > or =3% with 90% sensitivity and 81% specificity. Age, sex, body mass index, smoking history, atopy, and the presence of asthma did not affect the FeNO level and its correlation with ISE. The NIOX-MINO (R) and NObreath(R) agree with each other to a high degree. Both devices showed close correlation with ISE with similar cut-off value in identifying ISE > or =3%.