Treatment of Thoracolumbar Spinal Fracture Accompanied by Diffuse Idiopathic Skeletal Hyperostosis Using Transdiscal Screws for Diffuse Idiopathic Skeletal Hyperostosis: Preliminary Results
- Author:
Hisanori IKUMA
1
;
Shinichiro TAKAO
;
Yoichi INOUE
;
Tomohiko HIROSE
;
Keitaro MATSUKAWA
;
Keisuke KAWASAKI
Author Information
- Publication Type:Clinical Study
- From:Asian Spine Journal 2021;15(3):340-348
- CountryRepublic of Korea
- Language:English
-
Abstract:
Study Design This retrospective case series enrolled 13 patients who underwent posterior fixation with both transdiscal screws for diffuse idiopathic skeletal hyperostosis (TSDs) and pedicle screws (PSs) to treat spinal injury accompanied by diffuse idiopathic skeletal hyperostosis (DISH).
Purpose To describe the usefulness, feasibility, and biomechanics of TSD.
Overview of Literature Vertebral bodies accompanied by DISH generally have lower bone mineral density than normal vertebral bodies because of the stress shielding effect. This phenomenon tends to makes screw fixation challenging. To our knowledge, solutions for this issue have not previously been reported.
Methods Patients were assessed using the data on surgical time, estimated intraoperative blood loss, mean number of stabilized intervertebral segments, number of screws used, perioperative complications, union rate, and the three-level EuroQol five-dimensional questionnaire (EQ5D-3L) score at the final follow-up. The Hounsfield unit (HU) values of the screw trajectory area, and the actual intraoperative screw insertion torque of TSDs and PSs were also analyzed and compared.
Results The surgical time and estimated intraoperative blood loss were 165.9±45.5 minutes and 71.0±53.4 mL, respectively. The mean number of stabilized intervertebral segments was 4.6±1.0. The number of screws used was 4.9±1.3 for TSDs and 3.0±1.4 for PSs. One death occurred after surgery. The union rate and EQ5D-3L scores were 100% and 0.608±0.128, respectively. The HU value and actual intraoperative screw insertion torque of TSDs were significantly better than those of PSs (
p <0.001,p =0.033).Conclusions We were able to achieve stable surgical outcomes using the combination of TSDs and PSs. The HU value and actual intraoperative screw insertion torque were significantly higher for TSDs than for PSs. Based on these results, when treating thoracolumbar spinal fractures accompanied by DISH in elderly populations, the TSD could be a stronger anchor than the PS.