Fast Separation Method of Continuous Recording Data in Dynamic Electrocardiogram Database.
10.3969/j.issn.1671-7104.2021.04.020
- Author:
Xiaoling YANG
1
;
Ming HAN
1
;
Jin LI
1
Author Information
1. Shaanxi Medical Device Quality Testing Institute, Xi'an, 712046.
- Publication Type:Journal Article
- Keywords:
comparison;
database;
dynamic electrocardiogram;
record;
separate
- MeSH:
Electrocardiography;
Heart Rate;
Humans
- From:
Chinese Journal of Medical Instrumentation
2021;45(4):442-445
- CountryChina
- Language:Chinese
-
Abstract:
The square waves and pulses with obvious porter characteristics are inserted into each different adjacent existing dynamic electrocardiogram data to solve the time-consuming problem in the currently used manual input verification method. The standard database files are converted into analog siginals, then output automatically to the ECG acquisition device according to the sequence of the database files in one time. The data recorded in the acquisition equipment is separated according to the interval data protocol, so as to achieve the purpose of rapid digital examination of dynamic electrocardiogram.