The clinical application of magnetic resonance-guided radiotherapy
10.3760/cma.j.cn113030-20200317-00118
- VernacularTitle:MR引导的放疗系统临床应用
- Author:
Biaoshui LIU
;
Xuan GUO
;
Shouliang DING
;
Bin WANG
;
Yongbao LI
;
Yunfei XIA
;
Yi OUYANG
;
Xiaoyan HUANG
;
Chengguang LIN
- From:
Chinese Journal of Radiation Oncology
2021;30(2):134-139
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To investigate the clinical feasibility of the Unity radiotherapy system guided by magnetic resonance imaging.Methods:Twenty-four patients were enrolled and received a total of 384 fractions of treatment at Unity system. According to the treatment site, all patients were divided into head-neck, abdomen-thorax, pelvic, spine and limb groups. The patients were set-up without external laser. And then, the time required at different stages in online treatment process and the registration error of each fraction were separately calculated. The geometric deformations of MR images were weekly measured by using MR geometric deformation phantom. At last, the Arccheck was used to perform the dose verification of reference plan, online plan and offline plan.Results:The mean duration of radiotherapy in the five groups were 29.1, 27.6, 26.6, 25.6 and 32.0 min, respectively. The set-up errors in the left-right, superior-inferior and anterior-posterior direction in the five groups were: head-neck group (0.08±0.06 cm, 0.16±0.13 cm, 0.08±0.05 cm), abdomen-thorax group (0.23±0.18 cm, 0.50±0.47 cm, 0.12±0.1 cm), pelvic group (0.25±0.19 cm, 0.32±0.25 cm, 0.11±0.09 cm), spine group (0.46±0.38 cm, 0.26±0.26 cm, 0.13±0.07 cm) and limb group (0.33±0.30 cm, 0.34±0.23 cm, 0.08±0.06 cm), respectively. In the central region, the geometric deformation of MR was less than 0.3 mm, and that of the sphere with a diameter of 500 mm was less than 2.1 mm. The meanγ pass rate of the reference plan, online plan and offline plan were 97.92%, 97.84% and 94.58%, respectively.Conclusions:MR-guided radiotherapy has great potential for clinical application, whereas the process of Unity system is relatively complex. The synergy of different departments has a great impact on the treatment, which needs further optimization.