Differences in dosimetric parameters between 125I seed implantation with degradable catheters and free-hand
10.3760/cma.j.issn.0254-5098.2021.01.013
- VernacularTitle:携带 125I粒子导管和徒手粒子植入后剂量学差异的模体研究
- Author:
Yun LIU
;
Chuang HE
;
Qinghua LIANG
;
Yuxiao CHEN
;
Liangshan LI
;
Jing YUAN
;
Tingyuan LI
;
Xuequan HUANG
- From:
Chinese Journal of Radiological Medicine and Protection
2021;41(1):62-65
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To compare the differences in dosimetric parameters of target areas between 125I seed implantation with degradable catheters and by hand under the assistance of a real-time intraoperative treatment planning system (TPS). Methods:Forty-two simulated lesions were divided into a degradable catheter group and a free-hand group, with twenty-one lesions in each group. 125I seeds were implanted according to the TPS.The pre-plan and post-implant dosimetric parameters were collected, including the minimum dose ( Dmin), maximum dose ( Dmax), mean dose ( Dmean), conformal index (CI), external index (EI), homogeneity index (HI), minimum prescription dose delivered to 90% of the target volume ( D90), and the mean percentage of volume receiving 90% of the prescription doses ( V90). The Bland-Altman method was employed to analyze the consistency of pre-plan and post-implant dosimetric parameters and the Wilcoxon rank-sum test was used for the comparison of the two groups. Results:According to the Bland-Altman analysis, the dosimetric parameters of the two groups were all in agreement before and after seed implantation except for the Dmin and V90 of the free-hand group.Furthermore, the degradable catheter group had smaller error ranges of Dmax ( Z=-3.824, P<0.005), CI ( Z=-1.962, P<0.005), HI ( Z=-2.352, P<0.005), D90 ( Z=-2.453, P<0.005), and V90 ( Z=-3.159, P<0.005). Conclusions:The dosimetric parameters of 125I seed implantation with degradable catheters under the assistance of real-time TPS have good pre-plan and post-implant consistency and smaller error ranges.