Improvement of detection method for acetaldehyde, butyraldehyde and isobutyraldehyde in workplace air
10.11763/j.issn.2095-2619.2020.05.014
- Author:
Guanlin CHEN
1
;
Jiaheng HE
;
Yingying ZHAO
1
;
Chunxia LUO
1
;
Zhiheng XU
1
;
Shaoxin HUO
1
;
Songgen CHEN
1
Author Information
1. Foshan institute of occupational disease prevention and control Foshan, Guangdong 528000, China
- Publication Type:Journal Article
- Keywords:
Workplace;
Air;
Acetaldehyde;
Butyraldehyde;
Isobutyraldehyde;
Gas chromatography;
Solvent desorption
- From:
China Occupational Medicine
2020;47(05):578-581
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE: To improve the standard detection method for acetaldehyde, butyraldehyde and isobutyraldehyde in the air of workplace. METHODS: Acetaldehyde, butyraldehyde and isobutyraldehyde in the air of workplace were collected using silica gel tube, desorbed with 45.0% ethanol, separated by a capillary column and detected by a flame ionization detector. RESULTS: The linear range of this method for detecting acetaldehyde, butyraldehyde and isobutyraldehyde were 1.57-1 568.00, 1.60-1 600.00 and 1.59-1 588.00 mg/L, respectively. All the correlation coefficients were greater than 0.999. The detection limits were 0.52, 0.46 and 0.54 mg/L, respectively. The desorption efficiency was 91.0%-103.0%. The within-run relative standard deviation(RSD) was 0.7%-1.7%.The between-run RSD was 2.0%-3.7%. The samples could be stored for at least 10 days at room temperature. CONCLUSION: This method can be used for simultaneous detection of acetaldehyde, butyraldehyde and isobutyraldehyde in the air of workplace.