Detecting resorcinol in workplace air by gas chromatography with capillary column
10.11763/j.issn.2095-2619.2019.04.016
- Author:
Xiaojuan LI
1
;
Yingxia HE
1
;
Baoli ZHU
1
Author Information
1. Jiangsu Provincial Center for Disease Prevention and Control Nanjing, Jiangsu 210009, China
- Publication Type:Journal Article
- Keywords:
Resorcinol;
Capillary column;
Gas chromatography;
Workplace;
Air
- From:
China Occupational Medicine
2019;46(04):474-479
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE: To establish a method for detecting resorcinol in workplace air by gas chromatography with capillary column. METHODS: The resorcinol in workplace air was collected into muiti-hole absorbing tubes with distilled water and detected by capillary chromatographic column by direct injection. RESULTS: The good linear range of resorcinol was 1.7-200.0 mg/L. The correlation coefficient was 0.999 9. The detection limit was 0.5 mg/L and the lower limit of quantitation was 1.7 mg/L. The minimum detection concentration was 0.7 mg/m~3(sample volume was 7.5 L). The standard recovery rate was 98.5%-102.6%. The within-run relative standard deviation(RSD) was 0.7%-3.6% and the between-run RSD was 1.8%-5.7%. CONCLUSION: This method has high sensitivity, accuracy and can effectively remove interference, which is suitable for determination of resorcinol in workplace air.