Research of Methods to Reduce Alarm Fatigue of Monitoring System.
10.3969/j.issn.1671-7104.2020.06.003
- Author:
Mengxing LIU
1
;
Zehui SUN
2
;
Wenyu YE
2
;
Sanchao LIU
2
;
Xianliang HE
2
;
Cheng WANG
2
;
Ye LI
1
Author Information
1. Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, 518000.
2. Shenzhen Mindray Bio-Medical Electronics Co., Ltd., Shenzhen, 518057.
- Publication Type:Multicenter Study
- Keywords:
alarm fatigue;
critical care;
intelligent threshold;
multi-parameter fusion;
nursing management
- MeSH:
Alert Fatigue, Health Personnel/prevention & control*;
Arrhythmias, Cardiac/diagnosis*;
Clinical Alarms;
Humans;
Intensive Care Units;
Monitoring, Physiologic
- From:
Chinese Journal of Medical Instrumentation
2020;44(6):481-486
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE:In order to solve alarm fatigue, the algorithm optimization strategies were researched to reduce false and worthless alarms.
METHODS:A four-lead arrhythmia analysis algorithm, a multiparameter fusion analysis algorithm, an intelligent threshold reminder, a refractory period delay technique were proposed and tested with collected 28 679 alarms in multi-center study.
RESULTS:The sampling survey indicate that the 80.8% of arrhythmia false alarms were reduced by the four-lead analysis, the 55.9% of arrhythmia and pulse false alarms were reduced by the multi-parameter fusion analysis, the 28.0% and 29.8% of clinical worthless alarms were reduced by the intelligent threshold and refractory period delay techniques respectively. Finally, the total quantity of alarms decreased to 12 724.
CONCLUSIONS:To increase the dimensionality of parametric analysis and control the alarm limits and delay time are conducive to reduce alarm fatigue in intensive care units.