Study on the synchronization of biventricular beats with the control mode of left ventricular assist device.
10.7507/1001-5515.202001046
- Author:
Fangqun WANG
1
;
Yao ZHANG
1
;
Wanqian HE
1
;
Si CHEN
2
;
Teng JING
2
;
Zhihao ZHANG
1
Author Information
1. School of Electrical and Information Engineering, Jiangsu University, Zhenjiang, Jiangsu 212013, P.R.China.
2. School of Fluid Machinery and Technology, Jiangsu University, Zhenjiang, Jiangsu 212013, P.R.China.
- Publication Type:Journal Article
- Keywords:
counter-pulse mode;
left ventricular assist device;
right ventricle failure;
synchronization
- MeSH:
Heart Failure/therapy*;
Heart-Assist Devices;
Humans;
Systole;
Ventricular Dysfunction, Right;
Ventricular Function, Right
- From:
Journal of Biomedical Engineering
2021;38(1):72-79
- CountryChina
- Language:Chinese
-
Abstract:
Right ventricular (RV) failure has become a deadly complication of left ventricular assist device (LVAD) implantation, for which desynchrony in bi-ventricular pulse resulting from a LVAD is among the important factor. This paper investigated how different control modes affect the synchronization of pulse between LV (left ventricular) and RV by numerical method. The numerical results showed that the systolic duration between LV and RV did not significantly differ at baseline (LVAD off and cannula clamped) (48.52%