- Author:
Yunlong LI
1
;
Jian ZHANG
2
;
Yanqiu WEI
1
;
Xiaojuan JIA
3
;
Xiaoyan LI
4
;
Liming TAN
5
;
Wenjun LIU
1
;
Limin YANG
1
Author Information
- Publication Type:Journal Article
- Keywords: PCR; aerosol; nucleic acid contamination; nucleic acid detection
- MeSH: Laboratories; Nucleic Acids; Pathology, Molecular
- From: Chinese Journal of Biotechnology 2021;37(2):673-679
- CountryChina
- Language:Chinese
- Abstract: Nucleic acid detection technique has good sensitivity and specificity and is widely used in in vitro diagnosis, animal and plant commodity quarantine, forensic identification, and other fields. However, it is susceptible to carryover contamination during the operation and leads to false-positive results, which seriously affects the detection accuracy. Therefore, finding an effective solution to prevent and eliminate nucleic acid carryover contamination has become particularly urgent. This study compared several different methods for removing nucleic acid contamination and confirmed that sodium hypochlorite solution and PCRguard reagent could effectively eliminate nucleic acid carryover in the liquid and on surfaces of different materials. Besides, the combination of sodium hypochlorite solution and PCRguard can solve the nucleic acid aerosol contamination. This study proposes solutions for the routine prevention of carryover contamination and removal of aerosol that has occurred in molecular diagnostic laboratories.