Assessment of Image Quality for Selective Intracoronary Contrast-Injected CT Angiography in a Hybrid Angio-CT System: A Feasibility Study in Swine
10.3349/ymj.2021.62.3.200
- Author:
Seongmin HA
1
;
Sunghee JUNG
;
Hyung-Bok PARK
;
Sanghoon SHIN
;
Reza ARSANJANI
;
Youngtaek HONG
;
Byoung Kwon LEE
;
Yeonggul JANG
;
Byunghwan JEON
;
Se-Il PARK
;
Hackjoon SHIM
;
Hyuk-Jae CHANG
Author Information
1. Graduate School of Biomedical Engineering, Yonsei University College of Medicine, Seoul, Korea
- Publication Type:Original Article
- From:Yonsei Medical Journal
2021;62(3):200-208
- CountryRepublic of Korea
- Language:English
-
Abstract:
Purpose:To compare image quality in selective intracoronary contrast-injected computed tomography angiography (SelectiveCTA) with that in conventional intravenous contrast-injected CTA (IV-CTA).
Materials and Methods:Six pigs (35 to 40 kg) underwent both IV-CTA using an intravenous injection (60 mL) and Selective-CTA using an intracoronary injection (20 mL) through a guide-wire during/after percutaneous coronary intervention. Images of the common coronary artery were acquired. Scans were performed using a combined machine comprising an invasive coronary angiography suite and a 320-channel multi-slice CT scanner. Quantitative image quality parameters of CT attenuation, image noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), mean lumen diameter (MLD), and mean lumen area (MLA) were measured and compared. Qualitative analysis was performed using intraclass correlation coefficient (ICC), which was calculated for analysis of interobserver agreement.
Results:Quantitative image quality, determined by assessing the uniformity of CT attenuation (399.06 vs. 330.21, p<0.001), image noise (24.93 vs. 18.43, p<0.001), SNR (16.43 vs. 18.52, p=0.005), and CNR (11.56 vs. 13.46, p=0.002), differed significantly between IV-CTA and Selective-CTA. MLD and MLA showed no significant difference overall (2.38 vs. 2.44, p=0.068, 4.72 vs. 4.95, p=0.078).The density of contrast agent was significantly lower for selective-CTA (13.13 mg/mL) than for IV-CTA (400 mg/mL). Agreement between observers was acceptable (ICC=0.79±0.08).
Conclusion:Our feasibility study in swine showed that compared to IV-CTA, Selective-CTA provides better image quality and requires less iodine contrast medium.