A study on the application of organ dose modulation technique to reduce breast radiation dose in chest CT imaging
10.3760/cma.j.cn112149-20190708-00262
- VernacularTitle:器官剂量调制技术在胸部CT中降低乳腺辐射剂量的研究
- Author:
Yongxian ZHANG
1
;
Yantao NIU
;
Lili ZHANG
;
Senlin GUO
;
Dandan LIU
;
Binbin YU
;
Jianxing WU
;
Tianliang KANG
;
Shijun WANG
;
Wei LI
Author Information
1. 首都医科大学附属北京同仁医院放射科 100730
- From:
Chinese Journal of Radiology
2020;54(6):587-591
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To investigate the effect of organ dose modulation (ODM) technique on reducing the breast radiation dose in chest CT scanning.Methods:In the phantom test, the PBU-2 adult chest module was used. The clinical chest scan protocol was used and three sets of scans performed on the chest module: (1) ODM off group, ODM was not used; (2) ODM part group, ODM was applied only in the breast region; (3) ODM all group, ODM was applied in the whole scan scope. Other scan parameters were same for the three groups, with smart mA applied. The volume CT dose index (CTDI vol) was recorded for all three groups. A long rod ionization chamber was placed in a fixed position in front of the right breast area to measure the breast skin dose (D). The contrast noise ratio (CNR) and the figure of merit (FOM) were measured respectively. In clinical research, 72 female patients who underwent chest CT scanning in Beijing Tongren Hospital Capital Medical University from August to November 2018 were retrospectively recruited. According to the application of ODM, the patients were divided into ODM off group (without ODM, 36 cases) and ODM part group (ODM applied in the breast region, 36 cases). The CTDI vol and the dose length product (DLP) were recorded. CNR, noise of images were measured and calculated, respectively. The image quality was evaluated by subjective evaluation scores. The one way ANOVA analysis was used in comparing the difference of CNR among the 3 groups in module test. As for clinical cases, the independent samples t test was used to compare the difference in CTDI vol, DLP, CNR and the noise between two groups; and the rank-sum test was used for comparison in image quality subjective evaluation. Results:In module test, the radiation dose was highest in ODM off group, and lowest in ODM all group. The CTDI vol were (6.90±0.02), (6.26±0.02) and (5.99±0.02) mGy, and the D values were (9.17±1.01), (8.01±0.92) and (7.58±0.87) mGy for ODM off group, ODM part group and ODM all group respectively. The CNR values of images with soft tissue algorithm reconstruction were highest in ODM off group and lowest in ODM all group, while no statistically significant difference was displayed ( P>0.05). The CNR values of the images with lung algorithm reconstruction showed the same trend, with statistically significant difference among the three groups ( F=154.732, P=0.006). The FOM of the lung and soft tissue algorithm images was maximized when the ODM was partially applied. As for clinical cases, compared with ODM off group, the dose of ODM part group showed significantly decreased, with CTDI vol decreased by 16.12% ( t=2.604, P=0.011), and the DLP decreased by 16.85% ( t=3.293, P=0.002). No significant difference was found in CNR, noise and subjective score by two doctors between two groups ( P>0.05). Conclusion:The application of ODM in chest CT imaging can reduce the radiation dose of breast with simultaneously maintaining the image quality.