Development of HPLC-MS/MS method for determination of linezolid in human plasma
10.13220/j.cnki.jipr.2016.04.029
- Author:
Ming-Xing GUO
1
Author Information
1. Department of Pharmaceutics, The General Hospital of the PLA Rocket Force
- Publication Type:Journal Article
- Keywords:
HPLC-MS/MS;
Linezolid;
Plasma
- From:
Journal of International Pharmaceutical Research
2016;43(4):740-743
- CountryChina
- Language:Chinese
-
Abstract:
Objective: To determine the concentration of linezolid in human plasma by high performance liquid chromatagraphy-tandem mass spectrometry(HPLC-MS/MS) so as to provide evidence for the rational administration in clinic. Methods: The plasma was extracted by acetonitrile with telmisartan as internal standard(IS). The mobile phase was 1 mmol/L ammonium acetate with 0.1% formic acid-acetonitrile with gradient elution. Quantitative analysis of [M+H]+ ions was m/z 338.1→m/z 296.3 (linezolid) and m/z 515.3→m/z 276.3 (telmisartan,the IS). Results: Excellent linear relationship was obtained from 0.23 to 46 µg/ml with correlation coefficient value of 0.9993. The lowest limit of quantitation was 46 ng/ml, the extraction recovery was from 87% to 109%. The intra- and inter- day precisions (relative standard deviations) ranged from 2.5% to 8.6%. Conclusion: The method is specific, sensitive and simple and can be used to determine the concentration of linezolid in clinic.