Effect of the High-Pitch Mode in Dual-Source Computed Tomography on the Accuracy of Three-Dimensional Volumetry of Solid Pulmonary Nodules: A Phantom Study.
10.3348/kjr.2015.16.3.641
- Author:
Sung Ho HWANG
1
;
Yu Whan OH
;
Soo Youn HAM
;
Eun Young KANG
;
Ki Yeol LEE
Author Information
1. Department of Radiology, Korea University Anam Hospital, Seoul 136-705, Korea. yuwhan@kumc.or.kr
- Publication Type:Original Article
- Keywords:
Pulmonary nodule;
Computed tomography;
Volumetric imaging, computer generated 3D;
Imaging phantom
- MeSH:
Humans;
Imaging, Three-Dimensional/instrumentation/*methods;
Lung/radiography;
Lung Neoplasms/*radiography;
Multiple Pulmonary Nodules/*radiography;
Phantoms, Imaging;
Solitary Pulmonary Nodule/*radiography;
Tomography, X-Ray Computed/instrumentation/*methods
- From:Korean Journal of Radiology
2015;16(3):641-647
- CountryRepublic of Korea
- Language:English
-
Abstract:
OBJECTIVE: To evaluate the influence of high-pitch mode (HPM) in dual-source computed tomography (DSCT) on the accuracy of three-dimensional (3D) volumetry for solid pulmonary nodules. MATERIALS AND METHODS: A lung phantom implanted with 45 solid pulmonary nodules (n = 15 for each of 4-mm, 6-mm, and 8-mm in diameter) was scanned twice, first in conventional pitch mode (CPM) and then in HPM using DSCT. The relative percentage volume errors (RPEs) of 3D volumetry were compared between the HPM and CPM. In addition, the intermode volume variability (IVV) of 3D volumetry was calculated. RESULTS: In the measurement of the 6-mm and 8-mm nodules, there was no significant difference in RPE (p > 0.05, respectively) between the CPM and HPM (IVVs of 1.2 +/- 0.9%, and 1.7 +/- 1.5%, respectively). In the measurement of the 4-mm nodules, the mean RPE in the HPM (35.1 +/- 7.4%) was significantly greater (p < 0.01) than that in the CPM (18.4 +/- 5.3%), with an IVV of 13.1 +/- 6.6%. However, the IVVs were in an acceptable range (< 25%), regardless of nodule size. CONCLUSION: The accuracy of 3D volumetry with HPM for solid pulmonary nodule is comparable to that with CPM. However, the use of HPM may adversely affect the accuracy of 3D volumetry for smaller (< 5 mm in diameter) nodule.