Enhanced Visualization of the Cervical Vertebra during Intraoperative Fluoroscopy Using a Shoulder Traction Device
- Author:
Van Tri TRUONG
1
;
Fidaa AL-SHAKFA
;
Ghassan BOUBEZ
;
Daniel SHEDID
;
Sung-Joo YUH
;
Zhi WANG
Author Information
- Publication Type:Clinical Study
- From:Asian Spine Journal 2020;14(4):502-506
- CountryRepublic of Korea
- Language:English
-
Abstract:
Methods:In this study, a group of patients (n=33, device group) underwent surgery with the new device used to apply caudal traction to both shoulders, and another group of patients (n=33, matched control group) had surgery with the tape traction. Data about the lowest vertebra visible on lateral fluoroscopic view, installation time, skin irritation under the traction area, and postoperative brachial palsy were recorded, and these parameters were analyzed using the t-test.
Results:The mean numbers of visible cervical vertebra were 6.3±0.41 in the device group and 5.6±0.32 in the matched control group (p <0.01, unpaired t-test). The mean installation times were 83.9±5.15 minutes in the device group and 73.7±6.32 minutes in the matched control group (p <0.02). Seven patients from the matched control group presented with skin irritation. However, none of the patients from the device group had the condition (p =0.005, Pearson chi-square test). Postoperative brachial palsy was not observed in both groups.
Conclusions:The Cervision system is more effective and superior to tape traction in pulling the shoulders down to improve the visualization of the cervical vertebra on lateral fluoroscopic view during cervical spine surgery.