10.3760/cma.j.issn.1673-0860.2017.08.004
- VernacularTitle: 改良儿童声源定位能力评测系统的评价
- Author:
Jinsheng DAI
1
,
2
;
Jinfeng LIU
1
,
2
;
Juan ZHANG
1
,
2
;
Xiaohui WEN
1
,
2
;
Yanjun WANG
1
,
2
;
Zhiyong LIU
1
,
2
;
Jingyan DU
1
,
2
;
Ningyu WANG
1
,
2
Author Information
1. Department of Otorhinolaryngology Head and Neck Surgery, Beijing Chaoyang Hospital, Capital Medical University, College of Otolaryngology, Captial Medical University
2. Key Laboratory of Otolaryngology Head and Neck Surgery, Minisity of Education(Capital Medical University), Beijing 100020, China
- Publication Type:Journal Article
- Keywords:
Sound localization;
Children;
Program evaluation
- From:
Chinese Journal of Otorhinolaryngology Head and Neck Surgery
2017;52(8):580-585
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To evaluate the feasibility of modified test system for sound localization (SL) in children.
Methods:Modified system (6 male, 9 female) and traditional method ( 5 male, 5 female) were used for evaluation of minimum audible angle(MAA) and root-mean-square error(RMS) error of 4 to 6 years old children, and the results were compared to verify the accuracy and effectiveness of the modified test system for children sound localization.SPSS 17.0 software was used to analyze the data(t test).
Results:(1) Comparison of veracity of modified system and traditional test: when tested at the positive front position using modified system, MAA and RMS error were(3.23±1.00)° and (13.68±5.18)° respectively.When using traditional method, MAA and RMS error were(3.17±0.59)°and (13.96±4.56)° respectively. No statistical differences were found between two groups(t value was 0.16, -0.14, both P>0.05). (2) Comparison of time used were as followed: when using modified system, it was (14.67±1.95) min for MAA, and (6.67±1.35) min for RMS error. When using traditional method, it was (36.30±6.81) min for MAA, and (21.00±3.50) min for RMS error. Time used were significant shorter in modified system than in traditional method (t value was-9.78, -12.37, both P<0.05).
Conclusion:Modified test system for children sound localization is useful and reliable in children′s horizontal SL test.The time used of modified test system is shorter than that of traditional test system.