Impact of sealing method and layer number of autoclave bag on autoclave sterilization effect in the BSL-3 laboratory
10.3760/cma.j.issn.1003-9279.2019.01.007
- VernacularTitle: BSL-3实验室医疗废弃物专用包装袋的封口方式及层数对高压灭菌效果的影响
- Author:
Wei ZHEN
1
;
Kun CAI
;
Haiyue SHI
;
Jing ZHANG
;
Weifang HAN
;
Weimin ZHOU
;
Guizhen WU
Author Information
1. Biosafety levels 3 laboratory, National Institute for Viral Disease Control and Prevention, Chinese Center for Disease Control and Prevention, Beijing 102206, China
- Publication Type:Journal Article
- Keywords:
Biosafety level 3 laboratory;
Autoclave sterilization;
Autoclave bag
- From:
Chinese Journal of Experimental and Clinical Virology
2019;33(1):30-33
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To evaluate, the method of different packing layer and sealing mode for the packaging of the infectious wastes in order to achieve better autoclave sterilization effect.
Methods:In the autoclave cycles P4 and P9, different packing layer (single, double and triple) and sealing mode (slightly folded, autoclave tape, nylon tape closure) were applied to the internal infectious waste sterilization, biological indicators were used to detect the effect of autoclave sterilization of the upper, middle and lower layers, respectively.
Results:1) In the autoclave cycle P4, the triple autoclave bags packaging sealing method was based on the slight folding, autoclave tape, and nylon taple, the biological indicator negative rate of the internal infectious waste in the upper layer, the middle layer and the lower layer were respectively in the upper layer 100%, 0%, 0%, in the middle layer 67%, 20%, 0%, and in the lower layer was 0%. The double autoclave bags packaging sealing method was based on the slight folding, autoclave tape, and nylon taple, the biological indicator negative rate of the internal infectious waste in the upper layer, the middle layer and the lower layer were respectively the in the upper layer 100%, 85%, 0%, the middle layer 45%, 0%, 0%, and the lower layer 10%, 0%, 0%. the single autoclave bag packaging sealing method wa based on the slight folding, autoclave tape, and nylon taple, the biological indicator negative rate of the internal infectious waste in the upper, middle and lower layers were respectively the upper layer 100%, 100%, 0%, the middle layer 70%, 45%, 0%, and the lower layer 12.5%, 0%, 0%. (2) In the autoclave cycle P9, the triple autoclave bags packaging sealing method was based on the slight folding, autoclave tape, and nylon taple, the biological indicator negative rate of the internal infectious waste in the upper layer, middle and lower layers were respectively the upper layer all 100%, the middle layer 100%, 100%, 70%, and the lower layer 100%, 100%, 30%. The double autoclave bags packaging sealing method was based on the slight folding, autoclave tape, and nylon taple, the biological indicator negative rate of the internal infectious waste in the upper, middle and lower layers were respectively the upper layer all 100%, the middle layer all 100% and the lower layer is 100%, 100%, and 60%. The single autoclave bag packaging sealing method was based on the slight folding, autoclave tape, and nylon taple, the biological indicator negative rate of the internal infectious waste in the upper, middle and lower layers were all 100%.
Conclusions:When autoclave is recommended for sterilization of infectious waste, the best effect was achieved with a single unsealed autoclave bag. If it is necessary to seal, the autoclave bag should not be completely sealed, avoiding incomplete penetration of steam into the interior space, otherwise the infectious waste can not be completely sterilized.