Topology optimization on the configuration of titanium reconstruction plate for bridging mandibular angle defect
- VernacularTitle:下颌角缺损修复钛板构型的拓扑优化研究
- Author:
Zhi-xiong TANG
1
;
Ya-lan LI
1
;
Wei-peng GUO
1
;
Wen-tao JIANG
1
;
Yu-bo FAN
2
Author Information
1. Laboratory of Biomechanical Engineering, Sichuan University
2. Key Laboratory for Biomechanics and Mechanobiology of Ministry of Education, School of Biological Science and Medical Engineering, Beihang University
- Publication Type:Journal Article
- Keywords:
Mandibular angle defect;
Repair;
Titanium plate;
Configuration;
Topology optimization
- From:
Journal of Medical Biomechanics
2014;29(2):E167-E173
- CountryChina
- Language:Chinese
-
Abstract:
Objective To provide an optimization method and a basic configuration for the configuration design of mandibular defect-repair implant. Methods A topology optimization computation based on density method was adopted to make topology optimization on reconstruction titanium plate for repairing mandibular defect. Results The effects of volume fraction F and penalization factor p on the optimized configuration were compared and analyzed, respectively, and the optimum parameters were chosen to be F=0.65 and p=3.5. Using this set of parameters, optimized computation was performed on the mandibular defect-repair implant, and the three opening configuration was proved to be the basic optimization configuration. The result also showed that unilateral molar clenching had more significant effect on stress distributions in the implant as compared to incisal clenching. Conclusions A set of parameters suitable for optimizing configuration of mandibular defect-repair implant, as well as the basic configuration of optimized titanium plate were obtained, which could provide references for the design of mandibular defect-repair implant in clinic.