Research of TLD calibrated in different standard X-ray RQR radiation field in terms of eye lens dose H p(3)
- VernacularTitle: 不同标准X射线RQR辐射场刻度眼晶状体当量剂量
H p(3)比较研究 - Author:
Wenyan LI
1
;
Guiying ZHANG
1
;
Zhichao WANG
1
;
Zeqin GUO
1
;
Mengxiao KANG
1
;
Chuanye LIU
1
;
Wentao SHENG
1
;
Yue ZENG
1
;
Xi LUO
1
;
Linlin YUE
2
Author Information
- Publication Type:Journal Article
- Keywords:
Eye lens;
RQR field;
H p(3); Thermoluminescent dosimeter(TLD) - From: Chinese Journal of Radiological Medicine and Protection 2020;40(1):42-46
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To compare the calibration result of standard X-ray RQR radiation field between SSDL (NIRP) and CEA LIST LNHB(France), and to explore the feasibility of calibrating
H p(3) in standard X-ray RQR radiation field of SSDL(NIRP).
Methods:Using a column model with a diameter and high of 20 cm, TLD was calibrated in SSDL (NIRP) and CEA LIST LNHB (France) to measure the personal dose equivalent eye lens doseH p(3), X-ray RQR radiation field included RQR4(60 kV), RQR7(90 kV), RQR9(120 kV), with energy response, angle response and linear response.
Results:In terms of energy response, the calibration results of TLD in both SSDL (NIRP) and CEA LIST LNHB (France) were in good agreement. The difference between exposure value and response value was less than 10%. In terms of angle response, the calibration result of TLD in CEA LIST LNHB (France) was better in SSDL(NIRP). The difference between exposure value and response value in CEA LIST LNHB (France) was less than 6%, while the difference between exposure value and response value in SSDL(NIRP) was more than 10% at angle of 20°. In terms of linear response, both calibration result of SSDL (NIRP) and CEA LIST LNHB (France) were in good agreement.
Conclusions:The standard X-ray RQR field in SSDL (NIRP) can be used for the calibration ofH p(3).