Endoscopic ultrasonography-guided transgastric drainage of pancreatic fluid collections with a novel lumen-apposing metal stent
10.3760/cma.j.issn.1007-5232.2019.09.009
- VernacularTitle: 新型金属支架用于内镜超声引导下经胃胰周积液引流的临床研究
- Author:
Zhijie WANG
1
;
Qianqian MENG
;
Pingping ZHANG
;
Jie GAO
;
Jie CHEN
;
Lei WANG
;
Kaixuan WANG
;
Dong WANG
;
Zhendong JIN
;
Zhaoshen LI
Author Information
1. Department of Gastroenterology, Changhai Hospital, Naval Medical University, Shanghai 200433, China
- Publication Type:Journal Article
- Keywords:
Stents;
Endosonography;
Pancreatic fluicl collection;
Drainage
- From:
Chinese Journal of Digestive Endoscopy
2019;36(9):654-658
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To evaluate the efficacy and safety of a novel lumen-apposing metal stent(LAMS) in endoscopic ultrasonography(EUS)-guided transgastric drainage of pancreatic fluid collections(PFC).
Methods:A retrospective study of 46 patients was performed who underwent EUS-guided transgastric drainage of PFC with placement of LAMS from September 2015 to April 2017. Clinical data were reviewed and follow-up data were obtained by telephone and outpatient contact.
Results:A total of 49 LAMS were placed in 46 patients (2 LAMS were placed in 3 patient respectively for multiport access). The operation success rate was 95.9%(47/49), complications occurred in 14 patients(31.8%), including 2 severe complications (delayed hemorrhage and peritonitis). Additional intervention was performed in 10 patients(22.7%). The mean hospital stay was 6 days (1-40 days) and patients were followed for a mean time of 18.4 months (9-28 months). Treatment success was achieved in 40 patients(90.9%). Stent removal was performed after a mean time of 59.4 days (20-142 days), and recurrence rate was 10%(4/40) during the follow-up period.
Conclusion:Drainage of PFC using LAMS with special design is effective and relatively safe.