Quantitative evaluation and feasibility analysis of cone-beam computed tomography (CBCT) image quality based on three-dimensional parameters
10.3760/cma.j.issn.1004-4221.2019.10.019
- VernacularTitle: 基于三维参数定量评价CBCT图像质量方法及可行性分析
- Author:
Xinyu QI
1
;
Shipei LU
;
Xin YANG
;
Chengguang LIN
Author Information
1. State Key Laboratory of Oncology in South China, Collaborative Innovation Center for Cancer Medicine, Guangdong Key Laboratory of Nasopharyngeal Carcinoma Diagnosis and Therapy, Sun Yat-sen University Cancer Center, Guangzhou 510060, China
- Publication Type:Journal Article
- Keywords:
Tomography X-ray computed, cone-beam;
Evaluation method;
Feasibility
- From:
Chinese Journal of Radiation Oncology
2019;28(10):799-800
- CountryChina
- Language:Chinese
-
Abstract:
Cone-beam computed tomography (CBCT) image based on a single layer of image and the choice of regions of interest (ROI) is affected by individual subjectivity, making it difficult to compare the image quality. In this article, a quantitative evaluation method of CBCT image quality based on three-dimensional parameters was proposed, which was extended from single layer to multi-layer and from pixel to voxel. The selection range of ROI was discussed to reduce the error of artificial selection and realize the quantitative evaluation of image quality by three-dimensional parameters.