Development of the Extraction Reference Point Beam Diagnostic System for Proton Medical Accelerator.
10.3969/j.issn.1671-7104.2019.02.007
- Author:
Huan LIAN
1
;
Kaizhong DING
1
;
Yucheng WU
2
;
Yonghua CHEN
1
;
Junjun LI
1
;
Han LI
1
Author Information
1. Institute of Plasma Physics, Chinese Academy of Sciences, Hefei, 230031.
2. Hefei CAS Ion Medical and Technical Devices Co Ltd, Hefei, 230031.
- Publication Type:Journal Article
- Keywords:
Faraday cup;
LabVIEW;
PLC;
beam diagnostic;
scintillation screen;
signal acquisition and processing
- MeSH:
Proton Therapy;
instrumentation
- From:
Chinese Journal of Medical Instrumentation
2019;43(2):102-105
- CountryChina
- Language:Chinese
-
Abstract:
er to detect the beam quality of the SC200 superconducting cyclotron,measure the beam at the extraction reference and the acceptance of the accelerator is realized.This article mainly introduces the design that use the scintillation screen at the extraction reference to measure the beam profile,position and use the Faraday cup to measure the current intensity with 2.5 level accuracy.The remoted controlling of probes and the acquisition and processing of signal based on LabVIEW and PLC.