Clinical Experiences of Spinal Surgery using TSRH Instrument
10.4055/jkoa.1994.29.4.1170
- Author:
Jae Do KANG
;
Kwang Yeul KIM
;
Dong Gil KWAK
;
Jae Bon PARK
- Publication Type:Original Article
- Keywords:
TSRH(Texas Scottish Rite Hospital
- MeSH:
Constriction;
Female;
Follow-Up Studies;
Humans;
Male;
Methods;
Spinal Stenosis
- From:The Journal of the Korean Orthopaedic Association
1994;29(4):1170-1179
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
The main goal of spinal surgery using implant is a rigid fixation to provide the stability until solid fusion will occur. Recently, various implant fixation devices have been introduced and transpedicular screw fixation is the usual method. In the past, we obtained the implant-related complications like screw failure and rod breakage after using the modified Harrington rod. However, we obtained good results after follow up over one year using TSRH instrument. We experienced spinal surgery using TSRH instrument in 42 cases since 1991 and followed from one year to 28 months with average 16 months. We analyzed the 28 cases and evaluated the implant-related problems. The results were as follows: 1. Among 28 patients, 10 patients were operated due to fracture and 9 patients operated due to spinal stenosis. 2. The male patients were 16 cases and the female were 12 cases. 3. The TSRH instruments provided the rigid fixation with three points clamping mechanism. 4. The cross-linking plate of TSRH was found to increase stiffness and strength. 5. There were no case of screw breakage. 6. In functional results by Kirkaldy-Willis; foriteria the excellent cases were 15, and the good were 2 cases.