Axisymmetric Contact Stress analysis of an Artificial Hip Joint of the Conical Fitting Type
10.4055/jkoa.1989.24.6.1665
- Author:
Yong San YOON
;
Young Yong KIM
;
Kye Lim LEE
;
Seung Key KIM
;
Young Hoon KANG
- Publication Type:Original Article
- Keywords:
Contact stress;
Finite element analysis
- MeSH:
Ceramics;
Finite Element Analysis;
Friction;
Hip Joint;
Hip
- From:The Journal of the Korean Orthopaedic Association
1989;24(6):1665-1672
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
The purpose of this study is to investigate the effect of the ball-cone contact interface variation on the ceramic ball stress which is the dominant factor of the design of the moldular artificial hip joint. A contact stress analysis problem is formulated using the optimization principle and finite element analysis technique on the ball-cone system of three dimensional axi-symmetrical model, Numerical procedure is employed to get the solution for varying angles of cones and angular mismatches between the ball and cone interfaces. There is a fair amount of friction in the interface and included in the analysis model.