Comparison of microwave thermosphere ablation and conventional microwave ablation in ex vivo bovine livers
10.3760/cma.j.issn.1004‐4477.2019.04.014
- VernacularTitle:真圆微波消融与传统微波消融效果对比的离体实验研究
- Author:
Wenying CHU
1
;
Dan LI
;
Lei LONG
;
Jianmin DING
;
Yan ZHOU
;
Xiang JING
Author Information
1. 天津医科大学三中心临床学院 300170
- Keywords:
Ultrasonography;
M icrowave ablation;
Heat sink effect;
Ex vivo
- From:
Chinese Journal of Ultrasonography
2019;28(4):353-358
- CountryChina
- Language:Chinese
-
Abstract:
Objective To compare the effectiveness of microwave thermosphere ablation( M T A ) and traditional microwave ablation( M WA ) in ex v ivo bovine livers ,and to compare the degree of the heat sink effect in them .Methods T he non‐vessel model and vessel model were established using fresh bovine livers . In non‐vessel model ( n = 48 ) ,the same power‐time settings were used in both M T A group and MWA group ,w hich were 80 W 12 min ,90 W 10 min and 100 W 10 min ,respectively . Long‐axis diameter ( Dl) , short‐axis diameter( Ds1 ,Ds2) ,roundness index ( R) and the time of temperature rising to 60℃ at place of 10 mm from the needle were measured .In vessel model( n =144) ,different vessel diameters( 3 mm ,5 mm ,10 mm) and flow rates ( 15 cm/s ,20 cm/s ,30 cm/s) were setted . The maximum radius( Rmax) ,total area of ablation zone( Sz ) ,and the area difference ( Sdiff ) were analyzed ,the temperatures adjacent to the vessel were monitored simultaneously . Results In non‐vessel model ,with the same power and time settings ,the Dl of M T A group was significantly smaller than MWA group ( P < 0 .01 ) , however , there was no significant difference of Ds1 and Ds2 between the two groups( P >0 .05) . And M T A group created more spherical ablation zones ,since the R of M T A group were more close to 1 ( P <0 .01) . In M T A group ,the time of temperature rising to 60 ℃ at place of 10 mm from the needle was slower than MWA group ( P <0 .01) . In vessel model ,the Sdiff of M T A group were hardly affected by the vessel diameters and flow rates ( P >0 .05) ,and there was also no statistical significance among different flow rates( P >0 .05 ) in M WA group ,but the Sdiff of M WA group was significantly affected by the vessel diameters( P <0 .01) . And the Sdiff of M T A group was significantly smaller than M WA group when the vessel diameter was 10 mm ( P <0 .05) ,while there was no statistical significance between the two groups w hen the vessel diameter was 3 mm or 5 mm( P >0 .05) . Conclusions Compared to M WA ,M T A can produce sizable ,regular and more spherical ablation lesions with the same power and time ,meanwhile ,it is less affected by the heat sink effect .