Self-made minimally invasive pelvic channel instrument for treatment of pelvic ring fracture-dislocation
10.3760/cma.j.issn.1671-7600.2019.03.003
- VernacularTitle:自制骨盆微创通道器械辅助治疗骨盆环骨折脱位
- Author:
Tonglin CHEN
1
;
Weidou JIA
;
Lianjun YUE
;
Hongtao BAI
;
Xiaobo LI
;
Shuolei WANG
;
Xiangbin GUO
;
Ying XIAO
;
Yingkui ZHANG
Author Information
1. 北京朝阳急诊抢救中心骨一科 100122
- Keywords:
Pelvis;
Fractures,bone;
Fracture fixation,internal;
Internal fixator;
Surgical procedures,minimally invasive
- From:
Chinese Journal of Orthopaedic Trauma
2019;21(3):191-198
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the clinical efficacy of our self-made minimally invasive pelvic channel instrument in the treatment of pelvic ring fracture-dislocation.Methods A retrospective study was conducted of the 35 patients who had been treated for pelvic ring fracture-dislocation from December 2015 to November 2017 and fully followed up at Department I of Orthopaedis,Beijing Chaoyang Emergency Rescue Center.They were 25 males and 10 females,aged from 20 to 73 years (average,41.3 years).According to the Tile classification for pelvic fractures,there were 26 cases of type B (type B1 in 8,type B2 in 12 and type B3 in 6) and 9 cases of type C(type C1 in 5,type C2 in 3 and type C3 in 1).Infix or anterior plate combined with percutaneous internal fixation with sacroiliac cannulated screws was used in 11 patients,sacroiliac triangle fixation combined with percutaneous internal fixation with anterior ring cannulated screws in 5 ones,and internal fixation with cannulated screws for anterior and posterior rings in 19 ones.All the channels were established using our self-made minimally invasive pelvic channel instrument for internal fixation with cannulated screws.The time for each screw placement and the number of X-ray projection were recorded.Postoperative reduction,pelvic function at the final follows-ups and complications were recorded.Results A total of 84 cannulated screws were inserted in the 35 patients.The time for each cannulated screw placement ranged from 5 to 13 minutes (average,8.1 minutes);the number of X-ray projection for each screw placement ranged from 7 to 15 times (average,10.3 times).Postoperative CT showed that all the cannulated screws were located in the preset channels.According to the Matta radiological criteria,postoperative reduction quality was excellent in 29,good in 4 and fair in 2,giving a good to excellent rate of 94.3%.The follow-up time for the 35 patients ranged from 6 to 15 months (average,12.3 months).At the final follow-ups,all the patients showed fine fracture union.There was no loosening or breakage of screws in all but one patient in whom one sacroiliac screw became loose 3 months after surgery.By the Majeed criteria,the pelvic function was excellent in 27 cases,good in 5,fair in 3 and poor in one,giving a good to excellent rate of 91.4%.Conclusion Our self-made minimally invasive pelvic channel instrument can be used to establish pelvic channels,leading to more accurate placement,shorter operative time and less X-ray projection.