Performance Evaluation of the Stat Profile pHOx Ultra Blood Gas Analyzer
10.15263/jlmqa.2019.41.1.47
- Author:
Hye Young LEE
1
;
Sunyoung AHN
;
Hyun Ki KIM
;
Woochang LEE
;
Sail CHUN
;
Won Ki MIN
Author Information
1. Department of Laboratory Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea. wonkmin@gmail.com
- Publication Type:Brief Communication
- Keywords:
Stat Profile pHOx Ultra Blood Gas Analyzer;
Performance;
Evaluation
- MeSH:
Calcium;
Carbon Dioxide;
Critical Care;
Humans;
Hydrogen-Ion Concentration;
Indicators and Reagents;
Lactic Acid;
Magnesium;
Oxygen;
Partial Pressure
- From:Journal of Laboratory Medicine and Quality Assurance
2019;41(1):47-49
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
The objective of this study was to evaluate the analytical performance of the Stat Profile pHOx Ultra Blood Gas Analyzer (Nova Biomedical, USA), a new blood gas/chemistry analyser, including its precision and linearity, comparison studies, and the carry-over effect of commercial reagents and patient specimens. We assessed all the results on the basis of the Clinical and Laboratory Standards Institute guidelines. The following parameters were assessed: pH, partial pressure of carbon dioxide, partial pressure of oxygen, ionized calcium, ionized magnesium (iMg), and lactate concentration The total imprecision had a coefficient of variation of 0.0%–1.8%, and the linear measurement ranges for each parameter were all acceptable. In comparison with the Nova Critical Care Xpress Analyzer (Nova Biomedical, USA), the results indicated a good agreement, except for iMg. All carry-over ranges were between −0.5% and −1.4%. The Stat Profile pHOx Ultra Blood Gas Analyzer showed good analytical performance in terms of precision, linearity, comparison studies, and carry-over effect. The Stat Profile pHOx Ultra Blood Gas Analyzer can provide reliable measurements across a clinically relevant range and has potential use in laboratory tests.