Biomechanical effects of interbody cage height on cervical spine
10.16151/j.1004-7220.2017.03.003
- VernacularTitle:椎间融合器高度对颈椎生物力学影响
- Author:
Yuan-Jun ZHU
1
;
Zhong-Jun MO
;
Cheng-Fei DU
;
Yan-Bin ZHAO
;
Li-Zhen WANG
;
Yu-Bo FAN
Author Information
1. 北京航空航天大学生物与医学工程学院
- Keywords:
Interbody cage;
Intervertebral height;
Cervical curvature;
Biomechanics
- From:
Journal of Medical Biomechanics
2017;32(3):220-226
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the biomechanical effects of interbody cage height on cervical spine during anterior cervical discectomy and fusion (ACDF) surgery,so as to provide references for selection of interbody cage.Metheds The finite element model of normal cervical spine (C2-7) was built and validated,and the cages with different height (5,6,7,8 mm) were implanted into C5-6 disc (cage 5,6,7,8 model).All the models were loaded with pure moment of 1.5 N · m to produce flexion,extension,lateral bending and axial rotation motions on cervical spine,and the effects of cage height on range of motion (ROM),facet joint stress,intervertebral pressure in cervical spine were investigated.Results The intervertebral angle at the fusion segment increased by 0.68° with per 1 mm-increase in height.The ROM in C5-6 after cage implantation was smaller than 0.44°.The influence of cage height on ROM in C4-5 was greater than that in C6-7,and the changes of ROM in non-fusion segments were smaller than 7.3%.The cage height variation had a smaller impact on the facet joint stress and intervertebral pressure.The stresses in the capsular ligament,cage and screw-plate system increased gradually with the increase of cage height,and the stresses in cage 6,7,8 models were much higher than those in cage 5 model.Conclusions For patients who need implanting fusion cage,the cage height should be 0-1 mm greater than the original intervertebral space height.