A comparative study of intracerebral veins and tributaries by SWI with different slice thicknesses
10.3760/cma.j.issn.1008-1372.2018.12.017
- VernacularTitle:不同层厚SWI对大脑内静脉及属支成像的比较研究
- Author:
Guanfeng CHEN
1
;
Yijuan XIANG
;
Changlian TAN
;
Hongchun ZHANG
;
Bingqing SUN
Author Information
1. 福建医科大学附属泉州第一医院影像科
- Keywords:
Cerebral veins/RA;
Magnetic resonance imaging
- From:
Journal of Chinese Physician
2018;20(12):1824-1828
- CountryChina
- Language:Chinese
-
Abstract:
Objectives To investigate the imaging ability of magnetic sensitive weighted imaging (SWI) scans with 2.0 mm and 1.0 mm slice thickness on intracerebral veins and branches.Methods 47 healthy volunteers were recruited,including 20 males and 27 females,aged 21-45 (30.1 ± 12.93).Two SWI sequences with thickness of 2.0 mm and 1.0 mm were performed in all healthy volunteers.After reconstructed by minimal intensity projection (MIP),the display rate and diameter of intracerebral veins and branches were counted by two doctors.McNemar test and paired t-test were used for statistical analysis.Results (1) The display rates of bilateral internal cerebral veins,papular veins and left ventricular veins were 100% in both scanning schemes,while the display rates of left and right septal veins,anterior caudate nucleus veins and right ventricular veins on SWI sequence images with 2.0 mm thickness were 90.0% (left),91.0% (right) and 87.2% (right),91.5%,respectively.The display rates on SWI sequence images with 1.0 mm thickness were 97.9% (left),94.0% (right),97.9% (left),95.7% (right) and 95.7% respectively.(2) There were significant differences in the diameter of bilateral septum pellucidum veins and bilateral anterior caudate nucleus veins between the two scanning schemes (P < 0.05).While there was no significant difference in the diameter of bilateral internal cerebral veins,bilateral papular veins and bilateral intraventricular veins (P >0.05).(3) The displaying rate of veins with diameter >0.9 mm was 100% for both scanning schemes,while the displaying rate of veins with diameter ranging from 0.6-0.9 mm (including 0.6 mm and 0.9 mm) in the two scanning schemes had significant difference (P < 0.05).Conclusions SWI sequence can clearly show the anatomical shape of intracerebral veins and branches.In contrast to SWI of the scan slices thickness of 2.0 mm,the SWI of scan slices thickness of 1.0 mm showed more efficiency in display thinner brain venous.Therefore,using SWI sequence scanning scheme with 1.0 mm slice thickness will be more conducive to observe and study intracerebral veins and branches.