Clinical Utility and Cross-Reactivity of Insulin and C-Peptide Assays by the Lumipulse G1200 System.
10.3343/alm.2018.38.6.530
- Author:
Jongwon OH
1
;
Jae Hyeon KIM
;
Hyung Doo PARK
Author Information
1. Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea. nayadoo@hanmail.net
- Publication Type:Original Article
- Keywords:
C-peptide;
Diabetes mellitus;
Insulin;
Lumipulse G1200;
Performance
- MeSH:
C-Peptide*;
Diabetes Mellitus;
Humans;
Insulin*;
Methods;
Proinsulin
- From:Annals of Laboratory Medicine
2018;38(6):530-537
- CountryRepublic of Korea
- Language:English
-
Abstract:
BACKGROUND: Measurement of insulin and C-peptide concentrations is important for deciding whether insulin treatment is required in diabetic patients. We aimed to investigate the analytical performance of insulin and C-peptide assays using the Lumipulse G1200 system (Fujirebio Inc., Tokyo, Japan). METHODS: We examined the precision, linearity, and cross-reactivity of insulin and C-peptide using five insulin analogues and purified proinsulin. A method comparison was conducted between the Lumipulse G1200 and Roche E170 (Roche Diagnostics, Mannheim, Germany) systems in 200 diabetic patients on insulin treatment. Reference intervals for insulin and C-peptide concentrations were determined in 279 healthy individuals. RESULTS: For insulin and C-peptide assays, within-laboratory precision (% CV) was 3.78–4.14 and 2.89–3.35%, respectively. The linearity of the insulin assay in the range of 0–2,778 pmol/L was R2=0.9997, and that of the C-peptide assay in the range of 0–10 nmol/L was R2=0.9996. The correlation coefficient (r) between the Roche E170 and Lumipulse G1200 results was 0.943 (P < 0.001) for insulin and 0.996 (P < 0.001) for C-peptide. The mean differences in insulin and C-peptide between Lumipulse G1200 and the Roche E170 were 19.4 pmol/L and 0.2 nmol/L, respectively. None of the insulin analogues or proinsulin showed significant cross-reactivity with the Lumipulse G1200. Reference intervals of insulin and C-peptide were 7.64–70.14 pmol/L and 0.17–0.85 nmol/L, respectively. CONCLUSIONS: Insulin and C-peptide tests on the Lumipulse G1200 show adequate analytical performance and are expected to be acceptable for use in clinical areas.