Three Cases of Femoral Stem Failure in Rotating Hinge Revision Total Knee Arthroplasty: Causes and Surgical Considerations.
10.4055/cios.2018.10.2.260
- Author:
Nemandra A SANDIFORD
1
;
Jonathan RA PHILLIPS
;
Diane L BACK
;
Andrew D TOMS
Author Information
1. Complex Arthroplasty Unit, St George's Hospital, London, UK. nemsandiford@hotmail.com
- Publication Type:Case Report
- Keywords:
Fracture;
Stem;
Arthroplasty;
Knee;
Revision
- MeSH:
Arthroplasty;
Arthroplasty, Replacement, Knee*;
Hip;
Knee;
Prostheses and Implants
- From:Clinics in Orthopedic Surgery
2018;10(2):260-264
- CountryRepublic of Korea
- Language:English
-
Abstract:
Catastrophic failure of stemmed components in arthroplasty is an uncommon but a serious complication. Stem fractures and techniques for addressing these have been described following hip arthroplasty but much less so following total knee arthroplasty (TKA). We review three cases of catastrophic failure of the stem in rotating hinge revision TKA prostheses. We discuss the possible mechanism of failure and review the current literature addressing this topic. Metaphyseal support needs to be optimized in order to minimize load transfer to the stem and to the junction (and the risk of fracture) if a modular component is used. When constrained components are used, radiographs need to be carefully assessed for signs of proximal loosening. Nonmodular stems are also an option in this situation.