Self-Gated Late Gadolinium Enhancement at 7T to Image Rats with Reperfused Acute Myocardial Infarction.
10.3348/kjr.2018.19.2.247
- Author:
Lei WANG
1
;
Yushu CHEN
;
Bing ZHANG
;
Wei CHEN
;
Chunhua WANG
;
Li SONG
;
Ziqian XU
;
Jie ZHENG
;
Fabao GAO
Author Information
1. Molecular Imaging Center, West China Hospital of Sichuan University, Chengdu 610041, China.
- Publication Type:Original Article
- Keywords:
Self-gated;
Myocardial infarction;
Flip angle;
MRI;
Late gadolinium enhancement;
Animal studies;
7T
- MeSH:
Animals;
Electrocardiography;
Gadolinium*;
Magnetic Resonance Imaging;
Methods;
Myocardial Infarction*;
Rats*;
Respiration
- From:Korean Journal of Radiology
2018;19(2):247-255
- CountryRepublic of Korea
- Language:English
-
Abstract:
OBJECTIVE: A failed electrocardiography (ECG)-trigger often leads to a long acquisition time (TA) and deterioration in image quality. The purpose of this study was to evaluate and optimize the technique of self-gated (SG) cardiovascular magnetic resonance (CMR) for cardiac late gadolinium enhancement (LGE) imaging of rats with myocardial infarction/reperfusion. MATERIALS AND METHODS: Cardiovascular magnetic resonance images of 10 rats were obtained using SG-LGE or ECG with respiration double-gating (ECG-RESP-gating) method at 7T to compare differences in image interference and TA between the two methods. A variety of flip angles (FA: 10°–80°) and the number of repetitions (NR: 40, 80, 150, and 300) were investigated to determine optimal scan parameters of SG-LGE technique based on image quality score and contrast-to-noise ratio (CNR). RESULTS: Self-gated late gadolinium enhancement allowed successful scan in 10 (100%) rats. However, only 4 (40%) rats were successfully scanned with the ECG-RESP-gating method. TAs with SG-LGE varied depending on NR used (TA: 41, 82, 154, and 307 seconds, corresponding to NR of 40, 80, 150, and 300, respectively). For the ECG-RESP-gating method, the average TA was 220 seconds. For SG-LGE images, CNR (42.5 ± 5.5, 43.5 ± 7.5, 54 ± 9, 59.5 ± 8.5, 56 ± 13, 54 ± 8, and 41 ± 9) and image quality score (1.85 ± 0.75, 2.20 ± 0.83, 2.85 ± 0.37, 3.85 ± 0.52, 2.8 ± 0.51, 2.45 ± 0.76, and 1.95 ± 0.60) were achieved with different FAs (10°, 15°, 20°, 25°, 30°, 35°, and 40°, respectively). Optimal FAs of 20°–30° and NR of 80 were recommended. CONCLUSION: Self-gated technique can improve image quality of LGE without irregular ECG or respiration gating. Therefore, SG-LGE can be used an alternative method of ECG-RESP-gating.