The application of IFCC reference methods in methodological evaluation of glycated hemoglobin measurement systems
10.3760/cma.j.issn.1009-9158.2018.11.007
- VernacularTitle:IFCC参考方法在五种糖化血红蛋白检测系统评价中的应用
- Author:
Tianjiao ZHANG
1
;
Yungang PU
;
Huijuan ZHOU
;
Rong MA
;
Jiangtao ZHANG
;
Daguang WANG
;
Chuanbao LI
;
Miao LI
;
Yufei WANG
;
Chuanbao ZHANG
Author Information
1. 100730,北京医院国家老年医学研究中心 卫生部临床检验中心 北京市临床检验工程技术研究中心
- Keywords:
Hemoglobin A,glycosylated;
Reference values;
Reference standards;
Chromatography,high pressure liquid;
Chromatography,ion exchange;
Electrophoresis,capillary;
Evaluation studies
- From:
Chinese Journal of Laboratory Medicine
2018;41(11):821-826
- CountryChina
- Language:Chinese
-
Abstract:
Objective To perform a methodological evaluation study of 4 HPLC based systems and a capillary electrophoresis based system , with the International Federation of Clinical Chemistry and Laboratory Medicine ( IFCC) glycated hemoglobin reference method as a comparative method .Methods 40 hemolysis samples of variety concentrations were prepared .The samples were measured by IFCC reference method and 5 glycated hemoglobin testing systems , respectively and trueness verification was performed according to the Clinical &Laboratory Standards Institute (CLSI) guideline EP9-A3.Whole blood samples were used to test the systems'precision, linearity and analytical interferences .Results The average CV of IFCC reference method results of 40 hemolysis samples was 1.4%( range from 0.2%to 2.5%).No outlier was found in the results of the 5 testing system.The slopes ranged from 0.9902 to 1.0267, and intercepts from -0.1526 mmol/mol to +0.1512 mmol/mol, squared correlation coefficient from 0.9962-0.9971, biases at two medical decision level were less than 0.3%HbA1c.Within-laboratory precisions were less than 2% (NGSP unit).Bias between all test results and predicted results were less than 5%.High concentration of glucose showed certain interference to glycated Hemoglobin tests but had little influence in clinical practice.Conclusions The results of the 5 testing system are comparable to the IFCC reference method , the results of precision and linearity evaluation meet the requirements of related guidelines .