Effects of common hemoglobin variants on 4 ion exchange high performance liquid chromatography methods based HbA1c measurements
10.3760/cma.j.issn.1009-9158.2018.10.012
- VernacularTitle:常见异常血红蛋白对4种离子交换高效液相色谱法检测糖化血红蛋白的影响
- Author:
Anping XU
1
;
Weidong CHEN
;
Yu ZHOU
;
Yong XIA
;
Yongqiang WANG
;
Hao XUE
;
Mingyang LI
;
Jie LI
;
Ruoyang ZHENG
;
Ling JI
Author Information
1. 518036,北京大学深圳医院检验科
- Keywords:
Hemoglobin A1c;
Hemoglobin variants;
Capillary electrophoresis;
Ion exchange high performance liquid chromatography
- From:
Chinese Journal of Laboratory Medicine
2018;41(10):765-769
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate effects of 7 common hemoglobin variants on HbA 1c measurements using 4 ion exchange high performance liquid chromatography methods .Methods Ninety five samples with hemoglobin variants were collected from January 2017 to February 2018 during HbA1c measurements in laboratary medicine of peking university shenzhen hospital .Samples with 7 common hemoglobin variants were measured using Sebia Capillary 2 Flex Piercing, Bio-Rad D-10, Arkray HA8180V, Tosoh G8, and MQ6000 Plus, respectively.Effects of 7 common hemoglobin variants on HbA 1c measurements by the 4 methods were analyzed using Capillary 2 Flex Piercing as a comparative method .All statistical analyses were carried out using SPSS software version 19.0 .Mean bias were calculated for samples with hemoglobin variants , box plot was established to display bias distribution .Results Hb New York showed no interference on the 4 HPLC mechods although Hb New York could not be detected .D-10 could detect 6 Hb variants, and showed clinically significant interference for Hb J-Bangkok, Hb G-Coushatta, and Hb G-Taipei.HA-8180V fast mode yielded no HbA1c values for Hb J-Bangkok, Hb G-Coushatta, and Hb G-Taipei.Hb E, Hb Q-Thailand, and Hb G-Honolulu produced significant negative biases for HA-8180V.G8 standard mode could detect 1 Hb variant, and showed significant negative biases for six Hb variants .MQ6000 Plus could separate six Hb variants , only Hb G-Coushatta and Hb G-Taipei produced significant negative biases for the system . Conclusions Some common hemoglobin variants can interfere with HbA 1c determination by the most popular methods in South China , which may lead to erroneous HbA 1c values.