Animal model experiment on application of new magnetic anchoring and guidance system on endoscopic submucosal dissection(with video)
10.3760/cma.j.issn.1007-5232.2017.12.011
- VernacularTitle:新型磁力锚导引系统在内镜黏膜下剥离术中应用的动物实验研究(含视频)
- Author:
Xinyue WAN
1
;
Honggang YU
Author Information
1. 武汉大学人民医院消化内科
- Keywords:
Traction,Therapeutic endoscopy;
Endoscopic submucosal dissection;
Magnetic anchoring and guidance
- From:
Chinese Journal of Digestive Endoscopy
2017;34(12):897-899
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the safety and efficacy of a new magnetic anchoring and guidance system on endoscopic submucosal dissection(ESD)in pig models. Methods Two skilled operators performed ESD at the rectum part of 2 pigs with the help of a new magnetic anchoring and guidance system. The dissection time,dissection rate per unit time, en-bloc resection rate and complications were analyzed. Results Five simulated lesions underwent ESD with the help of the magnetic anchoring and guidance system. The dissection time was 26.6 min and the dissection rate was 0.38± 0.1 cm2/min. All the lesions were dissected totally. There was no perforation or delayed bleeding. Conclusion The new magnetic anchoring and guidance system is effective and safe for pig rectal ESD. It can effectively expose submucosal layer,and assist resection under the endoscopic observation.