Experimental study of dosimetry for CTLD-J4000 energy differential type personal dosimeter
10.3760/cma.j.issn.0254-5098.2018.05.010
- VernacularTitle:CTLD-J4000型能量鉴别式个人剂量计的剂量学实验研究
- Author:
Yanqiu DING
1
;
Wen GUO
;
Aiying HU
;
Shuxia HAO
Author Information
1. 中国疾病预防控制中心辐射防护与核安全医学所辐射防护与核应急中国疾病预防控制中心重点实验室
- Keywords:
Energy identification;
Dosimetric performance;
Dosimeter
- From:
Chinese Journal of Radiological Medicine and Protection
2018;38(5):374-378
- CountryChina
- Language:Chinese
-
Abstract:
Objective To understand the dosimetric performance of energy differential type dosimeter.Methods A differential type dosimeter (CTLD-J4000) was chosed to conduct a series of dosimetric performance tests,including linear standard curves,relative response to mean photon radiation energy and incidence angle.Results The linear correlation of CTLD-J4000 energy differential dosimeter was good,with the average photon radiation energy response deviation within ± 20% and the maximum incidence angle response deviation of 10%.The energy identification could be achieved according to the energy reference value provided by the manufacturer to a certain extent.Conclusions Further experiments are needed to find out the factors influencing measurement result from different dose values and incidence angles.