Dosimetric verification on intensity modulated radiotherapy in Jiangsu province
10.3760/cma.j.issn.0254-5098.2018.02.008
- VernacularTitle:江苏省调强放疗剂量学验证研究
- Author:
Jin WANG
1
;
Chunyong YANG
;
Yuanyuan ZHOU
;
Ningle YU
Author Information
1. 210009,南京 江苏省疾病预防控制中心
- Keywords:
Intensity modulated radiotherapy;
Dose verification;
Thermoluminescent dosimeter;
Film
- From:
Chinese Journal of Radiological Medicine and Protection
2018;38(2):121-124
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the dosimetric verification method for absorbed dose and twodimensional dose distribution generated through multi-leaf collimator in radiotherapy.Methods A total of 8 medical accelerators were selected in Jiangsu province.Under 6 MV X-rays with a 5 cm × 5 cm field,the thermoluminescent dosimeter (TLD) and EBT3 radiochromic film (RCF) were used to audit the values calculated by treatment planning system (TPS).RCF was used to verify two-dimensional dose distribution under the 6 MV MLC beams for intensity modulated radiation therapy (IMRT).Results The relative dose deviation was from-1.4% to 6.8% for TLD and from-1.8% to 7.8% for RCF,with 7 accelerators meeting with the IAEA deviation requirement of ± 5%.The passing rates of two-dimensional dose distribution were between 73.7% and 97.0%,with 7 accelerators greater than 90%.Conclusions TLD and RCF are feasible and suitable for large-scale verification of IMRT TPS absorbed dose and twodimensional dose distribution.