Comparison of 2-dimensional and 3-dimensional contrast-enhanced ultrasound for quantitative analysis in liver neoplasms
10.3760/cma.j.issn.1004-4477.2018.05.006
- VernacularTitle:二维与三维超声造影定量分析技术评估肝肿瘤的对比研究
- Author:
Jiaying CAO
1
;
Peili FAN
;
Yi DONG
;
Lijuan MAO
;
Wenping WANG
Author Information
1. 复旦大学附属中山医院超声科
- Keywords:
Contrast-enhanced ultrasonography;
Liver neoplasms;
Quantitative analysis
- From:
Chinese Journal of Ultrasonography
2018;27(5):397-400
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the correlation and consistency of dynamic three-dimensional contrast-enhanced ultrasonography ( 3D-CEUS ) and two-dimensional contrast-enhanced ultrasonography (2D-CEUS) in the time-intensity curve ( TIC) parameters of quantitative analysis in liver neoplasms. Methods Thirty patients were examined with 2D-CEUS and dynamic 3D-CEUS.The dynamic raw data were analyzed by quantitative analysis software. The regions of interest were manually selected around the entire lesion;TIC curves were generated and fitted respectively. Quality of fit ( QOF) was used to evaluate the degree of fit between the fitted curve and the initial ccoefficients of S,PI and AUC were 0.92,0.72 and urve. The corresponding 2D-CEUS and 3D-CEUS quantitative parameters of the same lesion were compared,including peak intensity ( PI),slope ( S),area under curve (AUC),time to peak (TTP) and mean transit time (MTT). The correlation of the intensity-related quantitative parameters and the consistency of the time-related quantitative parameters were analyzed between the two kinds of CEUS. Results The consistency of time related quantitative parameters,including MTT and TTP,were 0.97 and 0.96,and the correlation were 0.94 and 0.93. The correlation 0.65,respectively. Conclusions The quantitative results of 3D-CEUS are stable and reliable,and have fair concordance with 2D-CEUS. Dynamic 3D-CEUS can be used for clinical evaluation of the blood perfusion of liver neoplasms.