Influence of adaptive statistical iterative reconstruction-V on image quality and radiation dose of abdominal CT in phantom
10.13929/j.1003-3289.201704047
- VernacularTitle:多模型迭代重建算法对腹部体模CT扫描图像质量和辐射剂量的影响
- Author:
Yaru CHAI
1
;
Jingjing XING
;
Jianbo GAO
;
Peijie LYU
;
Ping HOU
;
Suya WANG
;
Jie LIU
Author Information
1. 郑州大学第一附属医院放射科
- Keywords:
Tomography,X-ray computed;
Iterative reconstruction;
Radiation dosage;
Image quality;
Phantom
- From:
Chinese Journal of Medical Imaging Technology
2018;34(1):118-122
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the influence of preset adaptive statistical iterative reconstruction-V (ASIR-V) techniques on image quality and radiation dose reduction of abdominal CT in phantom,and to investigate the optimal ASIR-V level.Methods Abdominal anthropomorphic phantom was scanned using Revolution CT,when noise index (NI) were set as 6,8,10,12 and 14,respectively.Then 0-100% ASIR-V and conventional scan was performed and 55 sets of images were obtained.CT value,noise,subjective score and radiation dose were recorded,and the optimal ASIR-V was obtained.Subjective scores of images in each group were compared using rank sum test,and CT value,noise and radiation dose were compared with one way ANOVA and paired t test.Results The image subjective score unchanged when NI was 6,8 or 10,slightly increased when NI was 12 and 14 with 0-40% ASIR-V,and decreased above 50% ASIR-V at all NI.When NI was 6,8 or 10,more than 70% ASIR-V image subjective score fell below 3 points.When NI was 12 or 14 group,more than 60% ASIR V subjective score fell below 3 points.The image quality score of conventional scan had no difference with 40% ASIR-V when NI was 6,8 or 10,respectively (P=0.626,0.915,0.514),and inferior to 40% ASIR-V when NI was 12 or 14 (P=0.041,0.036),while in all NI group,image quality score of conventional scan was superior to 60% ASIR-V (P=0.021,0.012,0.015,0.014,0.007).CT values and image noises had no significant differences in different parts in all NI groups (all P>0.05).CT dose index volume (CTDIvol) continuing decreased with ASIR-V.Compared with that of conventional scan,at 40%,50% and 60% ASIR-V,CTDIvol reduced by 49.82%,62.51% and 71.63%,respectively.Conclusion Preset ASIR-V can reduce radiation dose obviously while maintaining the overall image quality,and 40%-60% ASIR-V can be recommended for abdominal CT in clinical application.