Feasibility of DDD Pacemaker Implantation Under EnSite Systerm Guidance With Micro X-ray
10.3969/j.issn.1000-3614.2018.02.016
- VernacularTitle:三维标测系统指导微射线下行双腔起搏器植入术的可行性探讨
- Author:
Rong-Jun ZHANG
1
;
Ming-Feng SHAO
;
Shou-Wen ZHANG
;
Yan-Jin WEI
;
Xin-Hua LI
Author Information
1. 临沂市人民医院 心内科
- Keywords:
EnSite system;
X-ray;
Pacemaker;
artificial
- From:
Chinese Circulation Journal
2018;33(2):178-180
- CountryChina
- Language:Chinese
-
Abstract:
Objective: To explore the feasibility of DDD pacemaker implantation under EnSite system guidance with micro X-ray. Methods: DDD pacemaker implantation under EnSite system guidance with micro X-ray was performed in 5 special patients who needed cardiac pacing while should be avoid X-ray exposure. Results: All 5 patients were smoothly finished DDD pacemaker implantation, no wire dislocation or perforation occurred after the operation. The cumulative skin radiation dose was (2.20±0.45) mGy, operative X-ray exposure time was (11.60±2.07) s and the operation time was (58.40±4.04) min. Conclusion: DDD pacemaker implantation under EnSite system guidance with micro X-ray was safe and feasible.