Physical performance testing of Mobetron 2000 mobile intraoperative radiotherapy system
10.7687/j.issn1003-8868.2017.12.084
- VernacularTitle:Mobetron2000移动式术中放疗系统的物理性能测试
- Author:
Han ZHOU
1
;
Ze-Tian SHEN
;
Jun-Shu SHEN
;
Xi YUAN
;
Xi-Xu ZHU
;
Bing LI
Author Information
1. 南京总医院放疗科
- Keywords:
radiotherapy system;
physical performance;
radiotherapy;
accelerator;
dose
- From:
Chinese Medical Equipment Journal
2017;38(12):84-87
- CountryChina
- Language:Chinese
-
Abstract:
Objective To execute physical performance testing of Mobetron 2000 mobile intraoperative radiotherapy system.Methods Mobetron 2000 mobile intraoperative radiotherapy system had its indexes measured according to TG51 protocol such as percent depth dose (PDD),output dose,beam energy,field flatness and symmetry,X-ray pollution,dose linearity and repeatability.Three-dimensional water phantom was used for verification and acceptance when the measurement was executed,while the calimator end was tangent to the water surface and the nominal ource-skin distance was 50 cm.Results All the indexes of the system met the desired requirements except 12 MeV beam energy.Conclusion The system can be promoted clinically for safe,reliable and efficient treatment.