Determination of residual chloroform in medical polylactic acid membrane by gas chromatography
10.3969/j.issn.1001-3733.2018.01.006
- VernacularTitle:气相色谱法测定医用聚乳酸薄膜中三氯甲烷残留量
- Author:
Lihua LV
1
;
Hua HUANG
;
Shiyin LU
;
Zhiheng SU
Author Information
1. 广西医科大学附属口腔医院
- Keywords:
Gas chromatography;
Chloroform;
Electron capture detector(ECD);
Polylactic acid membrane
- From:
Journal of Practical Stomatology
2018;34(1):30-32
- CountryChina
- Language:Chinese
-
Abstract:
Objective: To establish a method using gas chromatography for the determination of the residual chloroform in the medical polylactic acid membrane. Methods: The direct aqueous injection gas chromatographic method was established for the determination using RESTEK RTX capillary-column chromatography with the electron capture detector(ECD) at 270 ℃ and with the column at 85 ℃,maintained for 3 min. The split ratio of sampling was 10∶1. The injector temperature was 200 ℃. The high purity nitrogen was used as the carrier gas with the flow of 1. 0 ml /min. The injection volume was 1 μl. Results: The calibration curve showed a good linearity within the range of 25-2 000 ng /ml (r = 0. 999 9). The limit of detection was 1. 50 ng /ml,and the limit of quantitation was 4. 62 ng /ml. The average recovery rate was 98. 35%,RSD = 1. 98%. Conclusion: Gas chromatography is sensitive and accurate for the determination of the residual chloroform in the medical polylactic acid membrane.