Comparison of corneal curvature and astigmatism measurements with four kinds of instruments before cataract surgery
10.3980/j.issn.1672-5123.2018.5.34
- VernacularTitle:四种不同仪器测量白内障患者术前角膜曲率和散光的比较
- Author:
Ya-Ru DU
1
;
Xue-Xi LI
Author Information
1. 解放军第一八0医院眼科医院
- Keywords:
OPD - Scan Ⅲ;
corneal curvature;
astigmatism vector
- From:
International Eye Science
2018;18(5):904-908
- CountryChina
- Language:Chinese
-
Abstract:
AIM: To compare corneal curvature and astigmatism values identified with OPD - Scan Ⅲ, lOL Master, Pentacam and KR - 8900 in cataract patients and to evaluate the consistencies and differences. METHODS: In this prospective clinical trial, 45 cataract patients (86 eyes), received treatment at the Department of Ophthalmology in our hospital from April 2017 to June 2017, underwent corneal parameters measurements with four instruments: OPD-Scan Ⅲ, Pentacam, IOL Master, and KR-8900. Measurements included corneal curvature (K1, K2, Km) and astigmatism parameters. The results were assessed using the paired samples t tests, Pearson coefficient,and Biand-Aitman plots. RESULTS: The difference of K1value measured by OPD-ScanⅢ and IOL Master was significant (P<0.001). Significant differences were also found in K2and Km values between OPD - Scan Ⅲ and Pentacam or IOL Master(P<0.05),while others had noncomparable results (P>0 05). All biometry measurements revealed good liner correlation (r=0.408-0.980, P<0.001). The Biand-Aitman showed that the corneal curvature and astigmatism vectors examined by OPD-Scan Ⅲ and Aii other devices had non-comparable results,except for the comparison of K1, K2and Kmmeasured by OPD-Scan Ⅲand KR-8900.CONCLUSION: OPD-Scan Ⅲ and KR-8900 can be interchangeable for the measurements of corneal curvature, while the others are not directly interchangeable. The astigmatic results obtained from OPD-Scan Ⅲ are not completely interchangeable with KR-8900,IOL Master or Pentacam.