Evaluation of the New Automatic Mycob.T Stainer and Scanner for Detecting Acid-fast Bacilli in China.
- Author:
Qiang JU
1
;
Meng Xue LI
2
;
Gang CHEN
3
;
Heng Xue WANG
4
;
Qiao Mei SHI
3
;
Xing GE
5
;
Zhen DING
2
;
Qi WANG
2
;
Li Chun XU
1
Author Information
- Publication Type:Journal Article
- Keywords: Automatic; Scanner; Stainer; Tuberculosis
- From: Biomedical and Environmental Sciences 2018;31(8):572-578
- CountryChina
- Language:English
-
Abstract:
OBJECTIVETo validate the performance of Mycob. T Stainer and Scanner (MTSS) for detecting acid-fast bacilli (AFB).
METHODSA total of 3,816 sputum samples from 1,515 tuberculosis (TB) suspects were tested at the Anhui Provincial Chest Hospital and the Linyi People's Hospital from April-August, 2016. Each specimen was placed on two smear slides. One slide was stained by the ziehl-neelsen (ZN) method to be read by conventional microscopy (CM). The other slide was stained and scanned by MTSS. All specimens were decontaminated with 4% NaOH, and then inoculated into solid culture. The performance of MTSS was assessed.
RESULTSMTSS produced higher average positivity rate (27.96%) as compared with the CM (26.83%). The overall sensitivity and specificity of MTSS were 78.9% and 93.9%, respectively. The sensitivity and specificity of CM was 77.4% and 95.0%, respectively.
CONCLUSIONMTSS exhibited a favorable performance in the detection of AFB. It may be an alternative to CM for screening TB.