Research on Field Strength of Radiated Immunity Test for Active Implantable Medical Devices.
10.3969/j.issn.1671-7104.2018.03.004
- Author:
Hongyi YU
1
;
Weiqiang ZHANG
1
;
Zhong GAO
1
Author Information
1. Shanghai Testing and Inspection Institute for Medical Devices, Shanghai, 201318. ##Email#.
- Publication Type:Journal Article
- Keywords:
active implantable medical devices;
electromagnetic compatibility;
radiated immunity
- MeSH:
Electromagnetic Fields;
Electromagnetic Radiation;
Immunity;
Prostheses and Implants;
Radiation Dosage
- From:
Chinese Journal of Medical Instrumentation
2018;42(3):170-172
- CountryChina
- Language:Chinese
-
Abstract:
In order to compare the radiation immunity test level of ISO 14708 series standards and the domestic and international environmental radiation standards and then to ascertain whether the radiated immunity test level has reached the limit of the radiation strength in the relevant radiation environmental standards, this paper calculated the radiation field intensity and power density according to the radiated immunity test of ISO 14708 standards, and compared with the limit of ICNIRP 1998 and GB 8702-2014.