Comparison and Discussion of National/Military Standards Related to Flow Measurement of Medical Injection Pump.
10.3969/j.issn.1671-7104.2018.03.017
- Author:
Nan ZHANG
1
;
Juan ZHOU
1
;
Jinlai YU
1
;
Ziyu HUA
1
;
Yongxue LI
2
;
Jiangang WU
2
Author Information
1. The Medical Engineering & Maitenance Center, Chinese PLA General Hospital, Beijing, 100853. ##Email#.
2. Institute for Drug and Instrument Control, Health Division of Logistical Support Department of CMC, Beijing, 100071. ##Email#.
- Publication Type:Journal Article
- Keywords:
flow;
measurement;
medical injection pump;
standard
- MeSH:
Equipment and Supplies;
standards;
Infusion Pumps;
standards;
Injections;
Military Personnel;
Reference Standards
- From:
Chinese Journal of Medical Instrumentation
2018;42(3):219-221
- CountryChina
- Language:Chinese
-
Abstract:
Medical injection pump is a commonly used clinical equipment with high risk. Accurate detection of flow is an important aspect to ensure its reliable operation. In this paper, we carefully studied and analyzed the flow detection methods of three standards being used in medical injection pump detection in our country. The three standards were compared from the aspects of standard device, flow test point selection, length of test time and accuracy judgment. The advantages and disadvantages of these standards were analyzed and suggestions for improvement were put forward.