A Torso Simulator Design for Implantable Nerve Stimulator Test.
10.3969/j.issn.1671-7104.2018.04.010
- Author:
Rui ZHANG
1
;
Lu LIU
2
;
Hongyi YU
3
;
Yue ZHUO
3
;
Zhong GAO
3
Author Information
1. School of Information & Control Engineering, Xi'an University of Architecture & Technology, Xi'an, 710055.
2. Center for Medical Device Evaluation, CFDA, Beijing, 100081.
3. Shanghai Testing and Inspection Institute for Medical Devices, Shanghai, 201318.
- Publication Type:Journal Article
- Keywords:
design;
implantable;
neurostimulator;
torso simulator
- MeSH:
Electromagnetic Fields;
Electromagnetic Phenomena;
Implantable Neurostimulators;
standards;
Torso
- From:
Chinese Journal of Medical Instrumentation
2018;42(4):265-267
- CountryChina
- Language:Chinese
-
Abstract:
This paper introduces ISO 14708-3:2017, the new edition of the international standard for implantable neurostimulator, and emphasizes the new requirements in the clause of protection from RF electromagnetic interference. To meet this new requirements, this paper presents a design of torso simulator for the testing of implantable neurostimulator. The design includes volume conductor, electrodes and grids, which can simulate the actual operating environment of implantable neurostimulator in RF electromagnetic interference testing. The torso simulator is verified by performance in the last part of the paper.