- Author:
Nan GAO
1
;
Li-Qin HUANG
1
;
Rui WANG
1
;
Jun-Jie JIA
1
;
Shuo WU
1
;
Jing ZHANG
1
;
Hong-Wei GE
2
Author Information
- Publication Type:Journal Article
- MeSH: Blood Donors; Enzyme-Linked Immunosorbent Assay; Humans; Syphilis Serodiagnosis; Treponema pallidum
- From: Journal of Experimental Hematology 2018;26(3):905-910
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVETo evaluate the accuracy and precision of 2 kinds of anti-treponema pallidum (anti-TP) ELISA reagents in our laboratory for detecting the anti-TP in voluntary blood donors, so as to provide the data support for use of ELISA reagents after introduction of chemiluminescene immunoassay (CLIA).
METHODSThe route detection of anti-TP was performed by using 2 kinds of ELISA reagents, then 546 responsive positive samples detected by anti-TP ELISA were collected, and the infections status of samples confirmed by treponema pallidum particle agglutination (TPPA) test was identified. The confirmed results of responsive samples detected by 2 kinds of anti-TP ELISA reagents were compared, the accuracy of 2 kinds of anti-TP ELISA reagents was analyzed by drawing ROC and comparing area under curve (AUC), and precision of 2 kinds of anti-TP ELISA reagents was compared by statistical analysis of quality control data from 7.1 2016 to 6.30 2017.
RESULTSThere were no statistical difference in confirmed positive rate of responsive samples and weak positive samples between 2 kinds of anti-TP ELISA reagents. The responsive samples detected by 2 kinds of anti-TP ELISA reagents accounted for 85.53%(467/546) of all responsive samples, the positive rate confirmed by TPPA test was 82.87%. 44 responsive samples detected by anti-TP ELISA reagent A and 35 responsive samples detected by anti-TP ELISA reagent B were confirmed to be negative by TPPA test. Comparison of AUC showed that the accuracy of 2 kinds of anti-TP ELISA reagents was more high, the difference between 2 reagents was not statistically significant. The coefficient of variation (CV) of anti-TP ELISA reagent A and B was 14.98% and 18.04% respectively, which met the precision requirement of ELISA test.
CONCLUSIONThe accuracy and precision of 2 kinds of anti-TP ELISA reagents used in our laboratory are similar, and using any one of anti-TP ELISA reagents all can satisfy the requirements of blood screening.