Calibration and Radiation Survey of High Dose Rate Remote Afterloading System.
- Author:
Jeong Ok LEE
1
;
Jeong Ku KANG
;
Sun Rock MOON
Author Information
1. Department of Radiation Oncology, Wonkwang University, School of Medicine, Korea.
- Publication Type:Original Article
- Keywords:
Aquaplast;
Surface dose;
6MV X-ray;
Skin-sparing effect
- MeSH:
Calibration*;
Carboxymethylcellulose Sodium
- From:Journal of the Korean Society for Therapeutic Radiology
1995;13(1):101-112
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
High Dose Rate Remote Afterloading system was installed at Wonkwang Universi-ty Hospital in January 1994. In this report, the calibration of a Gammamed 12-i High Dose Rate Remote Afterloading system and the radiation survey around the facility after design and construct a shieding room are discussed. The radiation survey of the facility indicates that the use of ordinary concrete shielding of existing room will provide adequate shielding. Also, the methodologies for performing source calibra-tion are presented.